Circulation - Pacing (transvenous)
Emergency transvenous temporary pacing complications are common. In large hospitals this procedure is usually performed by cardiologists outside of the emergency department. Temporary pacing by emergency physicians may occasionally be necessary, but positive chronotropic drug infusions and transcutaneous pacing are preferred where possible.
Indications
Bradycardia (HR <40bpm)
and
Unstable (altered mental status, SBP <90mmHg, angina, pulmonary oedema)
and
Unresponsive or unsuitable to medical therapy or transcutaneous pacing
Contraindications (absolute in bold)
Mechanical prosthetic tricuspid valve
Severe hypothermia (risk of arrythmia)
Anticoagulation or recent thrombolysis
Alternatives
Treat precipitant (drugs, ischaemia, electrolyte abnormality)
Atropine 0.5mg intravenously, repeat after 3-5 minutes if necessary, up to a maximum of 3mg
Isoprenaline 2-10mcg/min IV, titrated according to clinical response (risk of fall in blood pressure)
Adrenaline 2-10mcg/min IV titrated according to clinical response
Transvenous cutaneous
Immediate permanent pacemaker
Informed consent
Medical emergency
Consent is not required if the patient lacks capacity or is unable to consent
Brief verbal discussion is recommended if the situation allows
Potential complications
Lead
Failure (dislodgement, disconnection)
Heart
Dysrhythmia (asystole, VF, VT)
Perforation
Tamponade
Chest
Vascular damage, bleeding and haematoma
Pneumothorax or haemothorax
Air embolism
Diaphragm stimulation
Infection (local and systemic)
Thromboembolism
Procedural hygiene
Standard precautions
Surgical aseptic non-touch technique
PPE: surgical mask, protective eyewear, sterile gloves, sterile gown, sterile ultrasound cover and gel
Area
Resuscitation bay
Staff
Procedural clinician, nurse assistant
Additional clinician dedicated to monitoring
Equipment
Ultrasound
Pacing generator with new batteries
ECG machine or monitor with anterior leads attached
ECG adaptor to connect pacing catheter to monitoring anterior ECG electrode
Pulmonary artery catheter or introducer sheath (usually 6Fr) and sterile sleeve
Balloon tipped bipolar pacing catheter (usually 3-5Fr, leak excluded with inflation under sterile water)
3ml syringe (air-filled)
Positioning
Position as for right internal jugular central venous catheter
Supine, on an incline with head down 15 degrees
Head slightly rotated away from puncture site
Insertion site: Between medial and lateral heads of SCM muscle, lateral to the carotid, aiming to ipsilateral nipple
Medication
10ml lignocaine 1%
Consider analgesia and sedation
Sequence (insertion and connection of pacing wire)
Insert and secure a 6Fr introducer sheath via right internal jugular vein (see internal jugular vein central venous access guide)
Attach the positive terminal of the pacing generator to the positive (proximal) catheter terminal
Attach the V2 ECG lead to the negative (distal) catheter terminal using an ECG adaptor (any V lead will work)
This forms an exploring intracardiac ECG electrode to localise the lead tip
Attach sterile sleeve to pulmonary artery catheter sheath
Insert pacing wire through sheath to 15cm depth
Inflate floatation balloon with 1.5ml air, and lock catheter stopcock leaving syringe attached
Re-advance pacing wire, noting changes on V2 to ascertain correct placement of balloon
The P waves will increase in size as the catheter tip approaches the right atrium
The QRS complex will increase in size as the catheter tip approaches the right ventricle
If the QRS complex falls in size the catheter is in the inferior vena cava (pull back to 15cm, twist and re-advance)
If catheter-induced ectopic beats are seen, withdraw the catheter slightly then re-advance after ectopy ceases
ST segment elevation will occur when the catheter tip contacts the right endocardial wall (correct placement)
Subxiphoid ultrasound can be used to demonstrate the wire in the RV, and visualise mechanical capture
Secure pacing wire when ST segment elevation is present, noting this current depth (usually 35-45cm)
Deflate balloon by releasing the stopcock to the syringe (refilling syringe)
Remove negative (distal) catheter terminal connection from the V2 lead on the ECG
Attach negative (distal) catheter terminal to the negative terminal on the pacing box
Sequence (confirmation of capture and securing pacing catheter)
Set ventricular pacing at a rate of 80bpm, output of 10mA, high sensitivity of 0.8mA (VVI demand pacing)
Turn pacer on, and assess for electrical capture (QRS complex following each pacing spike on ECG)
Assess for mechanical capture by palpating a pulse equal to the pacemaker rate
Ask patient to cough and check wire does not dislodge (loss of pacing)
Coil excess pacing catheter and secure under a large sterile dressing
Sequence (setting threshold to ensure consistent capture)
Reduce output until capture lost (may be <1mA in the optimum position), this is the threshold
Increase output to two times threshold output (usually 2-3mA)
Sequence (setting pacemaker sensitivity)
Reduce to minimum sensitivity (asynchronous mode), and ensure complete capture
Adjust sensitivity to mid position (approximately 3mA)
Decrease rate until pacing is suppressed by intrinsic rhythm
Check sensing indicator signals each native beat
Pacer fails to sense - increase sensitivity
Pacer triggered by p or t waves (over-senses) - decrease sensitivity
Once sensitivity threshold determined, set millivoltage to half that value
Post-procedure care
Chest X-ray (confirm pacing wire in RV, exclude pneumothorax)
ECG to confirm a LBBB pattern (RBBB may indicate perforation or misplacement)
Bedside ultrasound to exclude pericardial fluid and pneumothorax
Documentation (completion, technique, attempts, guidewire removal, complications)
Tips
Prophylactic pacing of patients at high risk of AV block is unlikely to be indicated in the emergency department
‘Overdrive’ pacing of tachyarrhythmia is not generally recommended due to risk of rhythm deterioration
Pacing is generally unsuccessful for drug-induced bradycardia
Inserting a semirigid wire catheter without a balloon (same method) has higher complication rates
Defibrillation and cardioversion are safe in patients with temporary pacemakers
Discussion
Tachyarrhythmia caused by digoxin or QT prolongation (excluding atrial fibrillation), might benefit from overdrive pacing (pacing at a rate 40 bpm higher than the dysrhythmia for 10 seconds unsynchronised runs).
Overdrive pacing has been used effectively in hospital, but method is not well described or validated. There is also a risk of rhythm deterioration. We do not recommend its use in emergency department without specialist cardiology input.
Prophylactic pacing of patients at high risk of AV block is unlikely to be indicated in the emergency department. We recommend that such patients be monitored in the emergency department and paced only if required.
Peer review
This guideline has been reviewed and approved by the following expert groups:
Emergency Care Institute
Please direct feedback for this procedure to ACI-ECIs@health.nsw.gov.au.
References
Brignole M, Auricchio A, Baron-Esquivias G, et al. 2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Eur Heart J. 2013;34(29):2281-2329. doi:10.1093/eurheartj/eht150
Australian Resuscitation Council. Guideline 11.9 – managing acute dysrhythmias. East Melbourne: ARC; 2009. 11pp. Available from: https://resus.org.au/guidelines/
NSW Agency for Clinical Innovation. Transcutaneous pacing. Sydney: ACI; 2015. Available from:
https://www.aci.health.nsw.gov.au/__data/assets/pdf_file/0018/380313/Transcutaneous_Pacing.pdf
Roberts JR, Custalow CB, Thomsen TW. Roberts and Hedges’ clinical procedures in emergency medicine and acute care. 7th ed. Philadelphia, PA: Elsevier; 2019.
Dunn RJ, Borland M, O'Brien D (eds.). The emergency medicine manual. Online ed. Tennyson, SA: Venom Publishing; 2019.
Harrigan RA, Chan TC, Moonblatt S, Vilke GM, Ufberg JW. Temporary transvenous pacemaker placement in the Emergency Department. J Emerg Med. 2007;32(1):105-111. doi:10.1016/j.jemermed.2006.05.037
Gammage MD. Temporary cardiac pacing. Heart. 2000;83(6):715-720. doi:10.1136/heart.83.6.715
Ganz LI. Temporary cardiac pacing. In: UpToDate. Waltham (MA): UpToDate. 2019 Mar 1. Retrieved Apr 2019. Available from: https://www.uptodate.com/contents/temporary-cardiac-pacing
eTG complete. Melbourne: Therapeutic Guidelines; 2018 Mar. Bradyarrhythmias (retrieved April 2019). Available from: https://tgldcdp.tg.org.au/viewTopic?topicfile=bradyarrhythmias
Murray L, Little M, Pascu O, Hoggett KA. Toxicology handbook. 3rd ed. Sydney: Elsevier Australia; 2015.